Power Turn Elite™ Lateral Rotation Mattress System- Model 9800Lateral Rotation Therapy with True Low Air Loss |
|
The Power-Turn™ Elite Mattress System combines two clinically effective therapies (three clinically effective therapies when pulsation mode is added): Lateral Rotation and True Low Air Loss in a portable system. The mattress utilizes 18 independent 9″ deep air cells to provide proper pressure redistribution for the prevention and treatment of Stages I-IV pressure ulcers. Lateral Rotation Therapy reduces the risk of pulmonary complications for patients who are immobile. A Full, Left, or Right Turn mode is offered with either a 25º Small Turn or a 40º Big Turn. The mattress replacement is designed to fit all standard hospital beds and includes a high-density foam base, head bolster, and abduction wedge for additional patient safety. The Power-Turn™ Elite Lateral Rotation Mattress System supports patients weighing from 650 lbs. to 1000 lbs. The Power-Turn™ Elite meets California Technical Bulletin #117 for fire retardancy. An optional fire barrier is available to meet the most stringent fire safety requirements when requested for foam base. |
Bladders: |
Cover: |
|
Hose connections are color-coded for easy connection.
|
Blower Pump: The blower pump is controlled by a micro-processor which allows the caregiver total control of comfort and chosen therapy. Cycle time ranges from 5 to 95 minutes. The blower pump inflates the mattress within minutes. This reduces the waiting time for therapeutic treatment to begin.Blower Options: – Pulsating mode allows caregivers to provide pulsating low air loss therapy. – Automatic 2 position fowler sensor detects patient positioning and automatically increases power as needed in the sacral area. 10% adjustment for a 30º incline. 5% additional adjustment for a 40º incline. |
Head Rest and Leg Abductor: The headrest and leg abductor wedge provide support during the turning cycles. Head support is constructed from Viscotec™, a viscoelastic foam. The leg abductor is constructed from PPF™, Pressurized Polymer Foam. Both are covered with Vyvex-III™. |
MODEL 9800
|
Power-Turn Elite ™ – Lateral Rotation Therapy Mattress and Pump |
Dimensions:
|
36″ x 80″ x 9″ |
Weight:
|
23 lbs. |
Air Cells:
|
18 – 9 ” deep cells Ventilated low air loss bladders with air holes throughout the mattress. |
Side Bolster :
|
Patented air side rail bolsters are built into the mattress perimeter which supports the patient and ensures patient safety during rotation cycles. |
Mattress Cover:
|
Vyvex-III™ – Urethane coated, multi-stretch, low shear, moisture vapor permeable, quilted, removable, zippered cover, is anti-microbial, and meets Cal. Tech. #117 for fire-retardancy. |
Degree of Turn:
|
25º Small turn 40º Big turn |
WARRANTY
|
|
Mattress:
|
12 month limited |
Pump:
|
12 month limited |
Cover:
|
6 month limited |
ORDERING INFORMATION
|
|
Model 9800
|
36″ x 80″ x 9″ , 650 lbs capacity |
Model 9842
|
42″ x 80″ x 9″ , 800 lbs capacity |
Model 9848
|
48″ x 80″ x 9″ , 1000 lbs capacity |
Model 9800-AW
|
Abductor Wedge |
Model 9800-HR
|
Head Rest |
Model 9826
|
Replacement zippered cover |
Model 9826-42
|
42″ Replacement zippered cover |
Model 9826-48
|
48″ Replacement zippered cover |
Custom and bariatric sizes are available. Please state the patient’s weight when ordering
Use of side rails is recommended for patient safety. |
PDAC APPROVED
|
|
Medicare Code E0277
|
Power Turn™ Elite meets Group 2 Support Surface requirements for Medicare EO277. Patient must meet all medical criteria to qualify for reimbursement. |
MODEL 9801
|
Power-Turn Elite ™ Lateral Rotation Therapy- Pump only | |
Dimensions:
|
17″L x 7″W x 10.5″H | |
Weight:
|
12 lbs. | |
Electrical:
|
120V 60 hz | |
Listings:
|
UL® 544 Medical Equipment | |
Cycle Time:
|
5 – 95 Minutes | |
Air Flow:
|
Over 1800 Liters per minute | |
Current:
|
5A | |
Ground Resistance:
|
0.1 ohms maximum | |
Power Cord:
|
12 ft. cord with hospital grade plug | |
Modes of Operation :
|
Auto full turn Left turn Right turn Auto-firm Static |